1.3.18  Electrical noise
EN 61326-01, Class A
1.3.19  Line voltage
Universal 100 to 240 Vac ±10 percent, 50 to 60 Hz, no switches or jumpers
required, ¾-14 national pipe thread (NPT) conduit port
1.3.20  Pollution degree
2
1.3.21  Over voltage category
II
1.3.22  Relative humidity
5 to 95 percent (non-condensing)
1.3.23  Isolated output
Oxygen
4-20 mAdc, 950 ohm maximum with HART
®
or FOUNDATION
Fieldbus capability
only
Combustibles
4-20 mAdc, 950 ohm maximum
1.3.24 
Alarm
Alarm output relay - dry contact, form C, 30 mA, 30 Vdc capability
1.3.25 
Power consumption
750 W maximum
1.4  Ordering information
For ordering information, refer to the Rosemount OCX8800 Oxygen and
Combustibles Transmitter Product Data Sheet.