1.3.9  Calibration
Semi-automatic or automatic
1.3.10  Calibration gas mixtures recommended (reference
test gas bottles kit #1A9919G04)
0.4 percent O
2
, balance N
2
8 percent O
2
, balance N
2
1000 ppm CO, balance air
1.3.11  Calibration gas flow
7 scfh (3.3 L/m), regulated to 20 to 30 psig (1.38 to 2.07 barg)
1.3.12  Reference air
2 scfh (1 L/m), clean, dry, instrument-quality air (20.95 percent O
2
), regulated to 35
psig (2.41 barg)
1.3.13  Eductor air
5 scfh (2.5 L/m), clean, dry, instrument-quality air (20.95 percent O
2
), regulated to
35 psig (2.4 barg)
1.3.14 
Dilution air
0.1 scfh (0.05 L/m), clean, dry, instrument-quality air (20.95 percent O
2
), regulated
to 35 psig (2.41 barg)
1.3.15 
Blowback air (optional)
Clean, dry, instrument-quality air (20.95 percent O
2
), regulated to 60 psig (413
kPa) or greater and ambient temperature of 0 °F (-18 °C)
1.3.16 
Sensors housing
Type 4X, IP66 with fitting and pipe on reference exhaust port to clean, dry
atmosphere, two ¾-14 national pipe thread (NPT) conduit ports (when reference
air vents are routed to a dry area).
1.3.17 
Electronics housing
Type 4X, IP66 with fitting and pipe on reference exhaust port to clean, dry
atmosphere, two ¾-14 NPT conduit ports (when reference air vents are routed to
a dry area.
Description and specifications Manual
December 2022 00809-0500-4880
24 Emerson.com/Rosemount