3.2 Pre-programmed Regimens (PPRs)
The 300PV has 3 available Sport PPR’s or thirteen preprogrammed regimens, four
of which are intended to be user-defined. These PPR’s are the centerpiece of the
300 PV. They are designed for very specific clinical applications and are intended
to simplify the use of the device by allowing the clinician to educate the patient
on the use of the device with very little effort.
3.2.1 NMES PPR’s
3.2.1.1 PPR 1 — Summary
This is a sequenced 2-channel program for muscle re-education. All parameters
are preset. The clinician may adjust waveform, off time and treatment time as
desired.
Details
ADJUSTABLE PARAMETERS RANGE
Waveform Symmetric or Asymmetric
Treatment Time Unlimited, or 5-60 minutes
Off Time 0 to 60 seconds
PRESET PARAMETERS DEFAULT SETTINGS
Waveform Symmetric
Treatment Time (Timer) 30 minutes
Channel 1 Ramp-up time 3 seconds
Channel 1 ON time 12 seconds
Channel 1 Ramp-down time 2 seconds
Channel 2 Ramp-up time 2 seconds
Channel 2 ON time 9 seconds
Channel 2 Ramp-down time 1 second
Channel 2 starts 3 seconds after Channel 1
Off Time 30 seconds
Pulse Rate 35 Hz
Pulse Duration (Width) 300 s
3.2.1.2 PPR 2 — Summary
This PPR is a sequenced 2-channel program for muscle re-education. All
parameters are preset and adjustable with the exception of waveform and cycling,
which must be selected by the clinician.
19