Off Time 5 seconds
Pulse Rate 80 Hz
Pulse Duration (Width) 300 s
Channel Cycling Synchronous
3.2.1.7 PPR 7 — Summary
This regimen is a conventional TENS program for treatment of pain.
Details
ADJUSTABLE PARAMETER RANGE
Treatment Time Unlimited, or 5-60 minutes
PRESET PARAMETERS DEFAULT SETTINGS
Waveform Symmetric
Treatment Time 30 minutes
Modulation Rate 1 Hz
Amplitude Modulation 40% ( 40% decrease in intensity
at 1 Hz rate)
Pulse Rate 100 Hz
Pulse Duration 300 s
Channel Cycling Synchronous
3.2.1.8 CUSTOM USER 1 — Summary
Custom program with the therapeutic use defined by parameters set
by clinician.
Details
PARAMETER RANGE DEFAULT
Waveform Asymmetric or Symmetric Symmetric
Ramp Time 0 to 5 seconds 2 seconds
ON Time 0 to 60 seconds 10 seconds
OFF Time 0 to 60 seconds 30 seconds
Pulse Rate 2 to 150 Hz 35 Hz
Pulse Duration 50 to 400 s 300 s
Channel Cycling Asynchronous or Synchronous Synchronous
ON Time 1-60 seconds 10 seconds
23