22
3.2.1.5 PPR 5 — Summary
PPR 5 offers continuous stimulation unless used with an external hand or foot
remote switch (see 3.1.1 for details). As such, it is intended for use as a gait training
protocol. The intensity of each channel is independently adjustable. For the remote
hand switch, Channel 1 is active when the trigger switch is depressed, and Channel
2 is active when the trigger switch is released. For the foot switch, Channel 1 is
active when the weight of the foot is released from the foot switch, and Channel 2 is
active when the weight of the foot is applied to the foot switch.
Details
ADJUSTABLE PARAMETERS RANGE
Waveform Symmetric or Asymmetric
Treat me nt Time U nl im ited, or 5-6 0 min utes
PRESET PARAMETERS DEFAULT SETTINGS
Waveform Symmetric
Treat me nt Time (Ti me r) 30 min utes
Pulse Rate 35 Hz
Pulse Duration (Width) 300 s
3.2.1.6 PPR 6 — Summary
This regimen is a treatment for reduction of muscle spasms with a 2:1 ON/OFF
ratio.
Details
ADJUSTABLE PARAMETERS RANGE
Wavefor m Symmetric or Asymmetric
Treatment Time Unlimited, or 5-60 minutes
Off Time 0 to 60 seconds
Pulse Rate 2 to 150 Hz
Channel Cycling Synchronous or Asynchronous
PRESET PARAMETERS DEFAULT SETTINGS
Wavefor m Symmetric
Treat me nt Time 30 minutes
Channel 1 Ramp-up time 2 seconds
Channel 1 ON time 10 seconds
Channel 1 Ramp-down time 2 seconds
Channel 2 Ramp-up time 2 seconds
Channel 2 ON time 10 seconds
Channel 2 Ramp-down time 2 seconds