20
Details
ADJUSTABLE PARAMETERS RANGE
Waveform Symmetric or Asymmetric
Treatment Time Unlimited, or 5-60 minutes
Off Time 0 to 60 seconds
PRESET PARAMETERS DEFAULT SETTINGS
Waveform Symmetric
Treatment Time (Timer) 30 minutes
Channel 1 Ramp-up time 2 seconds
Channel 1 ON time 12 seconds
Channel 1 Ramp-down time 1 second
Channel 2 Ramp-up time 2 seconds
Channel 2 ON time 5 seconds
Channel 2 Ramp-down time 1 second
Channel 2 starts 5 seconds after Channel 1
Off time 30 seconds
Pulse Rate 45 Hz
Pulse Duration (Width) 300 s
3.2.1.3 PPR 3 --- Summary
This regimen is used for range of motion (ROM) or disuse atrophy arising from
traumatic injury and/or surgery.
Details
ADJUSTABLE PARAMETERS RANGE
Waveform Symmetric or Asymmetric
Treatment Time Unlimited, or 5-60 minutes
Off Time 0 to 60 seconds
Pulse Rate 2 to 150 Hz
Channel Cycling Synchronous or Asynchronous
PRESET PARAMETERS DEFAULT SETTINGS
Waveform Symmetric
Treatment Time (Timer) 30 minutes
Channel 1 Ramp-up time 2 seconds
Channel 1 ON time 10 seconds
Channel 1 Ramp-down time 2 seconds
00962.indd 2000962.indd 20 1/23/07 9:23:02 PM1/23/07 9:23:02 PM