3.5.5 Timer
Adjustable 5 to 60 minutes or unlimited
3.5.6 Remote Control Switch
Active regimen starts with closed contact
3.5.7 Low Voltage Indicator*
Indicator Threshold: 2.07 Volts (typical)
Shutdown Voltage: 1.77 Volts (typical) 1.50 Volt Minimum
*These voltages may be tested under NO load condition.
3.5.8 Physical Characteristics
Size 1.26" x 3.3" x 4.5"
Weight (with batteries) approximately 8 ounces
3.5.9 Environmental Conditions
Operational temperature 10° C to 40° C
Humidity (maximum) 75% RH
Transport and Storage -40° C to 70° C
Store in cool, dry place.
3.5.10 EN 60601-1 Classification
Type BF Applied Part
Internally powered only
Ordinary protection ingress of liquids
Continuous operation
Not suitable for use in the presence of a flammable anaesthetic
mixture with air or with oxygen or nitrous oxide