32
Ramp-up Time 0 seconds
ON Time 3 seconds
OFF Time 0 seconds
Pulse Rate 100 Hz/ 2 Hz
Pulse Duration 150 s
Channel Cycling Synchronous
3.3.7.3 Sports PPR 3 — Summary
For use in relaxation of muscle spasms.
Details
PRESET PARAMETERS DEFAULT SETTINGS
Waveform Symmetrical Biphasic
Treatment Time 30 minutes
Ramp-up Time 1 second
ON Time 5 seconds
OFF Time 1 second
Pulse Rate 10 Hz
Pulse Duration 300 s
Channel Cycling Synchronous
3.4 Accessories
Battery Type Four AA alkaline batteries- two in device at a time (part number
44800750) or rechargeable NiMH batteries (part number 200033-001).
Electrodes Any Empi electrode with a minimum active area of 5.1 cm
2
(roughly
1" x 1") – (Example: part number 86905220 - Ass'y, Electrode, 5220, 2"
x 2" Square, 4 Pack, (5000 Series), StimCare Carbon FM)
Lead wires Any Empi TENS/NMES lead wire (100 cm – 40”) with the touch
proof device connection (part number 193057-100).
Remote Switches
Use either Empi Foot Switch (part number 198887) or Empi Hand
Switch (part number 57202241).
3.5 Technical Data
3.5.1 Standard Measurement Conditions
Temperature 25°C ± 5°C
Load 1k1
Power Supply 3.0V DC
All tolerances ±10% unless noted.