3.2.1.9 CUSTOM USER 2 — Summary
Custom program with the therapeutic use defined by parameters set
by clinician.
Details
PARAMETER RANGE DEFAULT
Waveform Asymmetric or Symmetric Symmetric
Ramp Time 0 to 5 seconds 2 seconds
ON Time 0 to 60 seconds 10 seconds
OFF Time 0 to 60 seconds 30 seconds
Pulse Rate 2 to 150 Hz 35 Hz
Pulse Duration 50 to 400 s 300 s
Channel Cycling Asynchronous or Synchronous Synchronous
ON Time 1-60 seconds 10 seconds
3.2.2 HV PPR’s
3.2.2.1 HV PPR 1 — Summary
For use in increasing local circulation.
Details
ADUSTABLE PARAMETER RANGE
Treatment Time Unlimited, or 5-60 minutes
PRESET PARAMETERS DEFAULT SETTINGS
Treatment Time 30 minutes
Ramp-up Time 2 seconds
ON Time Continuous
Pulse Rate 100 Hz
Polarity Negative (red treatment wire is negative)
3.2.2.2 HV PPR 2 — Summary
For use in increasing local circulation.
Details
ADJUSTABLE PARAMETER RANGE
Treatment Time Unlimited, or 5-60 minutes