3.2.2.4 HV CUSTOM USER 2 — Summary
For use in increasing local circulation. This program is custom, with parameters
to be set by the clinician.
Details
PARAMETERS RANGE DEFAULT
Waveform Twin-peak pulse 100 s
Treatment Time Unlimited,
5-60 minutes 30 minutes
Ramp Time 2 seconds 2 seconds
ON Time 1-60 seconds 4 seconds
OFF Time 0-60 seconds 0 seconds
Pulse Rate 1-120 Hz 100 Hz
Pulse Duration 10 s 10 s
Polarity Pos/Neg/Reversing Negative
NOTE: For a more detailed explanation of the High Volt polarity selection, see
Figure 7 below.
Fig 7. High Volt Screen Display Options
3.3 Programming the Device
3.3.1 Setting Treatment Parameters
Programming the 300 PV is designed to be intuitive. The 300 PV has been
engineered to guide the clinician through the programming.
Step 1. Power ON the device by depressing the ON/OFF button
Step 2. Press the SETUP button
26
Screen Display
Negative
Positive
Neg/Pos
Pos/Neg
Instructions for Use
Red Treatment Lead Wire is
Negative
Red Treatment Lead Wire is
Positive
Reversing
Reversing
00962.indd 2600962.indd 26 1/23/07 9:23:03 PM1/23/07 9:23:03 PM