3.5.2 Outputs
Intensity:
NMES Asymmetrical biphasic 0 to 100mA ± 20%
NMES Symmetrical biphasic 0 to ± 100mA
TENS Symmetrical biphasic 0 to ± 50mA (40% decrease in intensity at 1Hz rate)
TENS IF Symmetrical biphasic 0 to ± 25mA
High Volt < 30 Volts ± 4V
30V to 300V ± 15%
Output Type:
NMES Constant current; compliance range 5001 to 10K1
HV Constant voltage; compliance range 5001 to 10K1
Pulse Waveforms:
NMES Asymmetrical and Symmetrical biphasic; interphase interval 800 sec
TENS Symmetrical biphasic, interphase interval <60 sec
HV Twin Peaks, 100 sec separation
Net DC: < 20 .A.
Pulse Width: (± 10% * 100 sec, ± 10 sec < 100sec)
NMES selectable from 50 to 400 sec
TENS fixed 300 sec
TENS IF fixed 125 sec
HV fixed 10 S @ 50% of amplitude w/ 1k1 load
Max RMS current: 21mA
Max Absolute Average current: 10 mA in 5001 load
Pulse Rate: NMES selectable from 2 to 150 pps
TENS fixed 100 pps
TENS IF modulated 1000 to 1150 pps
HV selectable from 1 to 120 pps
Max no load output voltage: NMES ± 140V
HV 400V
Accuracy of displayed values: Pulse Width, Pulse Rate and Intensity: < +/- 30%
3.5.3 Typical Waveforms
Standard measurement output voltage across purely resistive loads at maximum
intensity setting. Pulse Width and Vpp measured as shown across 500 ohm and
1k ohm loads.
33
00962.indd 3300962.indd 33 1/23/07 9:23:04 PM1/23/07 9:23:04 PM