77
No ultrasonic oscillations when the
footpedal is activated
Switch on the unit with the power switch
Checkiftheairowhandpiececordisinitsholder
Check the air connection of the unit
Check the connections between the unit, the handpiece cord and the handpiece
Check the connection of the footpedal
Change the handpiece
Send the unit to an approved EMS repair center
Decreasingorinsufcientultrasonic
power
Make sure that the instrument is properly positioned
Check the wear of the instrument and replace it if necessary
Change the handpiece
Send the handpiece to an approved EMS repair center
Air and/or powder is leaking from
the threads of the cap of powder
chamber
Check the seal and the cleanness on the thread of the powder chamber and of the cap
Replace the seal if necessary
Clean the cap and the powder chamber
Change the cap
Send the unit to an approved EMS repair center
Water leakage between handpiece
cord and handpiece
Check the O-ring on the handpiece cord
Send the handpiece cord to an approved EMS repair center
Leakage on the unit water supply Check the water supply connection
Change the water supply hose with connector
Send the unit to an approved EMS repair center