77
No ultrasonic oscillations when the
footpedal is activated
• Switch on the unit with the power switch
• Checkiftheairowhandpiececordisinitsholder
• Check the air connection of the unit
• Check the connections between the unit, the handpiece cord and the handpiece
• Check the connection of the footpedal
• Change the handpiece
• Send the unit to an approved EMS repair center
Decreasingorinsufcientultrasonic
power
• Make sure that the instrument is properly positioned
• Check the wear of the instrument and replace it if necessary
• Change the handpiece
• Send the handpiece to an approved EMS repair center
Air and/or powder is leaking from
the threads of the cap of powder
chamber
• Check the seal and the cleanness on the thread of the powder chamber and of the cap
• Replace the seal if necessary
• Clean the cap and the powder chamber
• Change the cap
• Send the unit to an approved EMS repair center
Water leakage between handpiece
cord and handpiece
• Check the O-ring on the handpiece cord
• Send the handpiece cord to an approved EMS repair center
Leakage on the unit water supply • Check the water supply connection
• Change the water supply hose with connector
• Send the unit to an approved EMS repair center