ENGLISH
FB-618/AU rev.E – ed.2019/08
19/48
PIEZON
®
and AIRFLOW
®
BOOST
Pressing hard on the center of the wireless pedal activates the BOOST mode and results in an
increase of power, as the following table shows:
AIRFLOW
®
Boost
Power Setting
0
1
2
3
4
5
6
7
8
9
10
Corresponding Boost Level
0
6
7
8
8
8
9
10
10
10
10
PIEZON
®
Boost
Power Setting
0
1
2
3
4
5
6
7
8
9
10
Corresponding Boost Level
0
6
7
8
9
10
10
10
10
10
10
Risk of tip breakage: use BOOST only with a tip suited for high power usage.
DO NOT use BOOST with ENDO files.
Wireless pedal battery saving
Each time the wireless pedal is released, it enters into a low power mode. Even if unused for long, it
is not required to remove the batteries.
To avoid an involuntary depletion of the wireless pedal batteries, in case the pedal remains pressed
without interruption for 10 minutes, it will automatically enter into switch-off mode.
To resume from the switch-off mode, it is required to first release the wireless pedal and then
power cycle the device (switch off for 30s and then power on again).
BOOST effect