Version PVT 800 :
Total length A: 1 190 mm (46.85 in)
Total weight: 15.54 kg (34.26 lb)
Measuring range length B: 800 mm (32 in)
Distance C: 41 mm (1.61 in)
Version PVT 1200 :
Total length A: 1 590 mm (62.60 in)
Total weight: 17.94 kg (39.55 lb)
Measuring range length B: 1 200 mm (47 in)
Distance C: 41 mm (1.61 in)
Version PVT 1600 :
Total length A: 1 990 mm (78.35 in)
Total weight: 20.14 kg (44.40 lb)
Measuring range length B: 1 600 mm (63 in)
Distance C: 41 mm (1.61 in)
Version PVT 2000 :
Total length A: 2 390 mm (94.09 in)
Total weight: 22.44 kg (49.47 lb)
Measuring range length B: 2 000 mm (79 in)
Distance C: 41 mm (1.61 in)
Version PVT 2400 :
Total length A: 2 790 mm (109.84 in)
Total weight: 24.74 kg (54.54 lb)
Measuring range length B: 2 400 mm (94 in)
Distance C: 41 mm (1.61 in)
Version PVT 3000 :
Total length A: 3 390 mm (133.46 in)
Total weight: 28.14 kg (62.04 lb)
Measuring range length B: 3 000 mm (118 in)
Distance C: 41 mm (1.61 in)
The weight data refer to the stainless steel housing versions. The aluminum housing
versions are 2.5 kg (5.51 lb) lighter.
The additional weight for small parts is: 1 kg (2.20 lb)
If using a collimator, pay attention to the documentation SD02822F.
5.1.3 Mounting requirements for level measurements
Conditions
The Gammapilot FMG50 is mounted vertically for level measurements.
To facilitate installation and commissioning, the Gammapilot FMG50 can be configured and
ordered with an additional support (order feature 620, option Q4: "Retaining bracket").