iTEMP TMT112 Commissioning
Endress+Hauser 21
Min indication Display the minimum process value. The process value is accepted at the beginning of the
measurement.
Min. process value will be changed to the actual process value on access. On reset to
factory default, the default value is entered.
+10000
Max indication Display the maximum process value. The process value is accepted at the beginning of the
measurement.
Max. process value will be changed to the actual process value on access. On reset to
factory default, the default value is entered.
-10000
Default values Entry: 182 (Reset to factory default settings)
0
Output simulation
Simulation mode
Entry of simulation mode.
Entry: Off
On
Off
Output current
Simulation value
Entry of simulation value (current).
Entry: 3.58 to 21.7 mA
Keycode
Security locking
Release code for setting up.
Entry: Lock = 0
Release = 281
281
Tag
Tag number
Entry and display of measurement point description (TAG).
Entry: 8 characters
-
Descriptor Entry and display of plant description.
Entry: 16 characters
-
Dev ID Display of device generation
Software rev Display of software version
e.g.: 11 indicates version 1.1
Product rev Display of unit version
e.g.: 1.0000 indicates version 1.00.00
1) Not for thermocouples (TC)
7.2.4 Supported HART® commands
No. Description Access
Universal Commands
00 Read unique identifier r
01 Read primary variable r
02 Read p.v. current and percent of range r