TMT84 Operation
Endress+Hauser 21
5Operation
5.1 Quick operation guide
!
Note!
Display and operating elements are only available locally if the head transmitter was ordered with a
display unit!
You have a number of options for configuring and commissioning the device:
1. Configuration programs Page 23
Both profile and device-specific parameters are configured exclusively via the fieldbus interface. You
can obtain special configuration and operating programs from various manufacturers for these pur-
poses.
2. Miniature switches (DIP switches) for diverse hardware settings, optional Page 24
You can make the following hardware settings for the PROFIBUS
®
PA interface using DIP switches
on the rear of the optional display:
• Entry of the device bus address
• Switching the hardware write protection on/off
• Switching (turning) the display 180 °
a0008323
Fig. 10: Head transmitter operating options
1 Configuration/operating programs for operation using PROFIBUS
®
PA (fieldbus functions, device parameters)
2 DIP switches for hardware settings on the rear of the optional display (write protection, device address, turn dis-
play)