Levelflex FMP51, FMP52, FMP54
Endress+Hauser 37
Differing from this, the following measuring error is present in the vicinity of the lower probe
end:
-80 (-3.15)
-60 (-2.36)
-40 (-1.57)
-20 (-0.79)
0
20 (0.79)
40 (1.57)
60 (2.36)
80 (3.15)
0
A
D
DC = 2
DC = 80
50 (1.97)
100 (3.94)
150 (5.91)
200 (7.87)
40 (1.57)
80 (3.15)
20 (0.79)
250 (9.84)
300 (11.8)
A0021480
13 Measuring error at the end-of-probe for rod and coax probes
A Distance from probe end [mm(in)]
D Measuring error: Sum of non-linearity, non-repeatability and hysteresis
-80 (-3.15)
-60 (-2.36)
-40 (-1.57)
-20 (-0.79)
0
20 (0.79)
40 (1.57)
60 (2.36)
80 (3.15)
0
DC = 80
A
D
50 (1.97)
100 (3.94)
150 (5.91)
200 (7.87)
20 (0.79)
250 (9.84)
300 (11.8)
A0021482
14 Measuring error at the end-of-probe for rope probes
A Distance from probe end
D Measuring error: Sum of non-linearity, non-repeatability and hysteresis