EasyManua.ls Logo

Endress+Hauser Levelflex FMP54 - Page 72

Endress+Hauser Levelflex FMP54
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Levelflex FMP51, FMP52, FMP54
72 Endress+Hauser
Temperature derating for FMP52
[°C] ([°F]) T
a
[°C] ([°F]) T
a
[°C]
([°F])
T
p
[°C]
([°F])
T
p
+200
(+392)
+200
(+392)
+82
(+180)
+79
(+174)
+74
(+165)
-40
(-40)
-40
(-40)
-50
(-58)
-50
(-58)
GT20: +60 (+140)
GT20: +56 (+133)
GT18: +53 (+127)
GT18: +51 (+124)
GT19: 38 (+100)
GT19: +38 (+100)
+80 (+176)
GT18/20: +79 (+174)
GT19: +74 (+165)
GT19: -36 (-33)
GT18: -37 (-35)
GT19: -36 (-33)
GT18: -37 (-35)
GT20: -38 (-36)
GT20: -38 (-36)
-40 (-40)
-40 (-40)
T
a
T
p
4 20 mA HART
A:
4 20 mA HART
4 20 mA
C:
90–253 VAC
K:
10.4–48 VDC
L:
[°C] ([°F]) T
a
[°C]
([°F])
T
p
+200
(+392)
+81
(+178)
-40
(-40)
-50
(-58)
GT20: +59 (+138)
GT18: +53 (+127)
+81 (+178)
GT18: -37 (-35)
GT20: -38 (-36)
-40 (-40)
[°C] ([°F]) T
a
[°C]
([°F])
T
p
+200
(+392)
+79
(+174)
-40
(-40)
-50
(-58)
GT20: +56 (+133)
GT18: +51 (+124)
+79 (+174)
GT18: -37 (-35)
GT20: -38 (-36)
-40 (-40)
PROFIBUS PA
FOUNDATION Fieldbus
Switch output
G :
2
PROFIBUS PA
FOUNDATION Fieldbus
G :
1
A0013633
GT18 = stainless steel housing
GT19 = plastic housing
GT20 = aluminum housing
A = 1 current output
C = 2 current outputs
G
1
, G
2
= PROFIBUS PA
1)
K, L = 4-wire
T
a
= ambient temperature
T
p
= temperature at the process connection
2)
1) For PROFIBUS PA and FOUNDATION Fieldbus the temperature derating depends on the usage of the
switch output. (G
1
: switch output not connected; G2: switch output connected).
2) For saturated steam applications the process temperature should not exceed 150 °C (302 °F). For higher
process temperatures use FMP54.

Table of Contents

Other manuals for Endress+Hauser Levelflex FMP54

Related product manuals