EasyManua.ls Logo

Endress+Hauser Levelflex FMP54 - Page 74

Endress+Hauser Levelflex FMP54
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Levelflex FMP51, FMP52, FMP54
74 Endress+Hauser
Temperature derating for FMP54 - version HT up to +450 °C (+842 °F )
[°C] ([°F]) T
a
[°C] ([°F]) T
a
[°C]
([°F])
T
p
[°C]
([°F])
T
p
+450
(+842)
+450
(+842)
+82
(+180)
+79
(+174)
+74
(+165)
-40
(-40)
-40
(-40)
-196
(-321)
-196
(-321)
GT20: +56 (+133)
GT20: +53 (+127)
GT18: +54 (+129)
GT18: +51 (+124)
GT19: +38 (+100)
GT19: +42 (+108)
+80 (+176)
GT18/20: +79 (+174)
GT19: +74 (+165)
GT18: -27 (-17)
GT18: -27 (-17)
GT19: -22 (-8)
GT19: -22 (-8)
GT20: -28 (-18)
GT20: -28 (-18)
-40 (-40)
-40 (-40)
T
a
T
p
4 20 mA HART
A:
4 20 mA HART
4 20 mA
C:
90–253 VAC
K:
10.4–48 VDC
L:
[°C] ([°F]) T
a
[°C]
([°F])
T
p
+450
(+842)
+81
(+178)
-40
(-40)
-196
(-321)
GT20: +55 (+131)
GT18: +53 (+128)
+81 (+178)
GT18: -27 (-16)
GT20: -28 (-18)
-40 (-40)
PROFIBUS PA
FOUNDATION Fieldbus
G :
1
[°C] ([°F]) T
a
[°C]
([°F])
T
p
+450
(+842)
+79
(+173)
-40
(-40)
-196
(-321)
GT20: +52 (+126)
GT18: +50 (+123)
+79 (+173)
GT18: -27 (-16)
GT20: -28 (-18)
-40 (-40)
PROFIBUS PA
FOUNDATION Fieldbus
G :
2
Switch output
A0013632
GT18 = stainless steel housing
GT19 = plastic housing
GT20 = aluminum housing
A = 1 current output
C = 2 current outputs
G
1
, G
2
= PROFIBUS PA
1)
K, L = 4-wire
T
a
= ambient temperature
T
p
= temperature at the process connection
1) For PROFIBUS PA and FOUNDATION Fieldbus the temperature derating depends on the usage of the
switch output. (G
1
: switch output not connected; G2: switch output connected).

Table of Contents

Other manuals for Endress+Hauser Levelflex FMP54

Related product manuals