EasyManua.ls Logo

Endress+Hauser Levelflex M FMP41C - Page 50

Endress+Hauser Levelflex M FMP41C
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning Levelflex M FMP41C with HART/4...20 mA
50 Endress + Hauser
Maximum measured error
Typical statements for reference conditions:
DIN EN 61298-2, percentage of the span.
If the reference conditions are not met, the offset/zero arising from the mounting situation may be
up to ±12 mm. This additional offset/zero can be compensated for by entering a correction
(function "offset" (057)) during commissioning.
Differing from this , the following measuring error is present in the vicinity of the probe end:
!
Note!
Please reenter the blocking distance in the function group "extended calibr." (05) function
"upper block.dist" (059) when installing the device in a high nozzle: upper blocking distance (UB)
= nozzle height (H) + 50 mm.
L00-FMP4xxxx-14-00-06-xx-001
Output: digital analogue
sum of non-linearity,
non-repeatability
and hysteresis
measurig range:
–up to 10 m: ±5mm
> 10 m: ± 0.05 %
for PA coated rope
measuring range:
up to 5 m: ±5 mm
> 5 m: ± 0.1 %
± 0.06 %
Offset / Zero ±4 mm ± 0.03 %
L00-FMP4xxxx-05-00-00-en-001 L00-FMP4xxxx-05-00-00-en-002
rod probe and coax probe
-80
-60
-40
-20
0
20
40
60
80
0 50 100 150 200 250 300
distance from probe end [mm]
sum of non-linearity, non-repeatability
and hysteresis [mm]
DC = 2
DC > 7
rope probe
-80
-60
-40
-20
0
20
40
60
80
0 50 100 150 200 250 300
distance from probe end [mm]
sum of non-linearity, non-repeatability
and hysteresis [mm]
DC > 7
UB
H
50

Other manuals for Endress+Hauser Levelflex M FMP41C

Related product manuals