Trouble-shooting Levelflex M FMP41C with HART/4...20 mA
68 Endress + Hauser
9.3 Application errors
Error Output Possible cause Remedy
A warning or alarm
has occurred.
Depending on the configuration See table of
error messages
( ä 66)
1. See table of error messages
( ä 66)
Measured value
(00) is incorrect
L00-FMP4xxxx-19-00-00-en-019
Measured distance
(008) OK?
yes
1. Check empty calibr. (005) and
full calibr. (006)
2. Check linearisation:
level/ullage (040)
max. scale (046)
diameter vessel (047)
Check table
no
An interference echo
may have been
evaluated.
yes
1. Carry out tank mapping
basic setup
No change off
measured value on
filling/emptying
L00-FMR2xxxx-19-00-00-en-014
Interference echo from
installations, nozzle
or extension on the
probe
1. Carry out tank mapping
basic setup
2. If necessary, clean probe
3. If necessary, select better
mounting position
E 641 (loss of
echo) after turn on
the power supply
If the instrument is configured to Hold by loss of
echo the output is set to any value/current.
noise level during the
initialisation phase to
high.
Repeat once more empty calibr.
(005).
Caution!
Before conformation change with
O
or
S
to the edit mode.
20 mA/100%
F m/ft
E m/ft
4 mA/0% t
D m/ft (008)
actual
expected
20 mA/100%
4 mA/0% t
actual
expected