Installation Levelflex M FMP40 and FMP45
20
FMP45: Installing with heat insulation
If process temperatures are high ( 200 °C), FMP45 must be included in normal tank insulation
to prevent the electronics heating up as a result of heat radiation or convection.
The insulation may not exceed beyond the points labelled "MAX" in the drawing.
Process connection with flange DN50...DN100
L00-FMP45xxx-17-00-00-en-001
Process connection with adapter G 1½" and 1½" NPT
L00-FMP45xxx-17-00-00-en-002
MAX
MAX
HT 400 °C
XT 280 °C
max. 400° C (752° F)max. 280° C (536° F)
tank insulation
tank insulation
high temperature version
(type of probe B)
extended temperature version
(type of probe A)
MAX
MAX
HT 400 °C
XT 280 °C
max. 400° C (752° F)max. 280° C (536° F)
tank insulation
tank insulation
high temperature version
(type of probe B)
extended temperature version
(type of probe A)