Levelflex M FMP40 and FMP45 Technical data
67
FMP45 (HT 400 °C)
If the temperature (T
2
) at the process connection is below -40 °C or above +80 °C, the permitted
ambient temperature (T
1
) is limited as shown in the following diagram (temperature derating):
L00-FMP45xxx-05-00-00-en-002
FMP45 (XT 280 °C)
If the temperature (T
2
) at the process connection is below -40 °C or above +80 °C, the permitted
ambient temperature (T
1
) is limited as shown in the following diagram (temperature derating):
L00-FMP45xxx-05-00-00-en-003
Storage temperature -40 °C … +80 °C (-40 °F … +176°F)
Climate class DIN EN 60068-2-38 (test Z/AD)
Vibration resistance DIN EN 60068-2-64 / IEC 68-2-64: 20…2000 Hz, 1 (m/s
2
)
2
/Hz
T
2
[°C]
T
1
[°C]
-200-250
0-100 -40
-60
-40
-33
-27
0
40
80
200 300
400
80
T
2
T
1
permitted process connection temperature
permitted ambient temperature
80 °C with separate
housing
54 °C compact:
F23 housing
66 °C
F12 housing
T12 housing
compact:
tank insulation
T
2
[°C]
T
1
[°C]
-200-250
0-100 -40
-60
-40
-29
-20
0
40
80
200 280
80
T
2
T
1
permitted process connection temperature
permitted ambient temperature
80 °C with separate
housing
56 °C compact:
F23 housing
66 °C
F12 housing
T12 housing
compact: