Technical data CM42-MAA101EBE00
38 Endress+Hauser
7Technical data
7.1 Input
7.1.1 Measured variables
Documentation of the connected sensor
7.1.2 Measuring range
Documentation of the connected sensor
7.1.3 Cable specification
7.2 Output
7.2.1 Output signal
2x 4 to 20 mA, passive, potentially isolated from one another
HART
7.2.2 Signal on alarm
3.6 to 21.5 mA (4.0 mA fixed for HART Multidrop mode)
7.2.3 Load
Max. load with supply voltage of 24 V: 500
Max. load with supply voltage of 30 V: 750
7.2.4 Current output, passive
Span
3.6 to 21.5 mA
CYK10, CYK20 with Memosens Max. cable length 100 m (330 ft)
Signal encoding FSK ± 0.5 mA via current signal
Data transmission rate 1200 Baud
Galvanic isolation Yes
Load (communication resistor) 250