Diagnostics, troubleshooting CM42-MAA101EBE00
24 Endress+Hauser
The following logbooks are available:
•Calibration logbook
– Log of calibrations and adjustments
– It is possible to call up the following details on every entry: time stamp, calibration method
and sensor type used
–Max. 15 entries
•Event logbook
– Log of the warnings and error messages
– It is possible to call up the following details on every entry: time stamp, event (diagnostic
message number), description, device status and event state
–Max. 50 entries
• Parameter logbook
– Log of changes in the configuration
– Every modified setting in the SETUP menu is recorded and displayed.
–Max. 50 entries
•User logbook
– Log of logins and logouts
– Each login and logout is recorded and displayed with a time stamp and the user name.
– The data are only recorded if user administration is enabled.
–Max. 50 entries
•Audit trail
– Chronological log of all logbook entries without the data logbook but with the device history
(see DIAG/Device information)
– Here, you can navigate to each individual logbook entry, regardless of the type of logbook,
and display the details.
– Max. 200 entries
You can also define and activate a data logbook:
• SETUP/General settings/Logbooks/Data logbook
• The data logbook consists of the recorded measured values in the scan rate you defined
including the associated time stamp.
• Max. 500 entries
3.8 Simulation, Resetting the measuring device
3.8.1 Simulation
You can output a simulated current value at the current output for test purposes.
The symbol for the simulation appears in the status bar of the display when the function is
switched on: