CM42-MAA101EBE00 Technical data
Endress+Hauser 41
Housing not sealed correctly, lack of strain relief
Dust can enter, cables can become loose, IP protection no longer guaranteed
‣ Make sure that cable glands cannot become loose and that the seals are installed close to
the housing.
‣ Once you have routed the cables through the glands, tighten the cable glands and the
associated nuts with a torque of 2 Nm to ensure the cables are secure.
‣ Pay attention to strain relief for the cables. Route the cables so that they are securely in
place.
‣ Make sure that the cable entries and glands are leak-tight.
7.4.3 Cable cross-section
Max. cable cross-section: 2.5 mm
2
(h 14 AWG), GND 4 mm
2
(h 12 AWG)
7.5 Performance characteristics
7.5.1 Current output response time
t
90
= max. 500 ms for an increase from 4 to 20 mA
7.5.2 Measured value resolution
→ Documentation of the connected sensor
7.5.3 Maximum measured error
→ Documentation of the connected sensor
7.5.4 Memosens maximum measured error
With digital data transmission, the measured value supplied by the sensor is relayed exactly at
the sensor input. The accuracy depends solely on the connected sensor and the quality of the
sensor adjustment.
7.5.5 Tolerance of current outputs
7.5.6 Repeatability
→ Documentation of the connected sensor
Current outputs, additional 25 μA
Current outputs, additional 25 μA