Liquiphant M FTL50(H), FTL51(H)
Endress+Hauser 35
Certificates and approvals
General approvals For Liquiphant M FTL50H, FTL51H the following approvals are available:
• EHEDG: Certification (from TNO, The Netherlands), Report No. V99.394
• 3A: 3A Certificate (USA), Authorization No. 459
!
Note!
For CIP (Clean in Place) and SIP (Sterilize in Place) processes the pressure and temperture specifications of the
process connections must be observed.
#
Warning!
To avoid risk of contamination, install according to the "Hygienic Equipment Design Criteria (HDC)" as stated
in the Subgroup Design Principles of the EHEDG, Doc. 8, July 1993. The flow of liquid during cleaning is
important and should be in compliance with HDC.
Other certificates See "Ordering information"
Combination
of housings and
electronic inserts
Based on the various certificates, permissible combinations of housings *) and electronic inserts are given in
the following table.
*) Abbreviations: Polyester = PBT, Steel 316L = St., Aluminium = Alu
Aluminium housing with separate connection compartment = Alu/sep
Process connections Order Code
L00-FTL5xxxx-01-05-xx-xx-001
Flush-mounted G ¾, G1 (with welding neck) GQ2, GW2 X
Tri-Clamp 2" (special seal from Hyjoin Limited, UK) TE2 X
Threaded pipe joint MA2, MC2, ME2 X
Flush-mounted (fork can be positioned) EE2 X
Aseptic HE2 X
DRD PE2 X
SMS UE2 X
Varivent
®
WE2 X
+
Certificate, applications Housing Electronic inserts
A Without any special certificate
(for non-hazardous area)
PBT, St., Alu, Alu/sep. FEL51/52/54,
FEL55/56/57/58/50A
D Overfill protection to WHG (Germany) PBT, St., Alu, Alu/sep. FEL51/52/54,
FEL55/56/57/58/50A
B ATEX II 3G EEx nC IIC T6, WHG PBT, St., Alu, Alu/sep. FEL54
ATEX II 3G EEx nC IIC T6, WHG
ATEX II 3D T85°C
St., Alu, Alu/sep. FEL54
C ATEX II 3G EEx nA II T6, WHG PBT, St., Alu, Alu/sep. FEL51/52,
FEL55/56/57/58/50A
ATEX II 3G EEx nA II T6, WHG
ATEX II 3D T85°C
St., Alu, Alu/sep. FEL51/52,
FEL55/56/57/58/50A
E ATEX II 1/2 G, EEx de, WHG Alu/sep. FEL51/52/54,
FEL55/56/57/58/50A
F ATEX II 1/2 G, EEx ia IIC T6, WHG PBT, St., Alu, Alu/sep. FEL55/56/57/58/50A
ATEX II 1/2 G, EEx ia IIC T6, WHG
ATEX II 1/2 D, T80°C
St., Alu, Alu/sep. FEL55/56/57/58/50A