Liquiphant M FTL50(H), FTL51(H)
Endress+Hauser 21
Operating conditions
Installation Installation instructions
!
Note!
The switchpoints of the Liquiphant M are at other positions to those of the previous version Liquiphant II.
Examples of mounting Examples of mounting with regard to the viscosity ν of the liquid and the amount of build-up
Optimum mounting, without problem even with high viscosity:
Switch points on the sensor depend on the mounting position, with reference to water,
density 1 g/cm
3
, 23 °C, p
e
0 bar.
L00-FTL5xxxx-06-05-xx-xx-001
Mounting from above Mounting from below Mounting from the side
13 mm
4mm
4mm
Position the fork so that the narrow edge of the tines is vertical.
This ensures that the liquid can run off easily.
L00-FTL5xxxx-11-05-xx-xx-001
Vertical from above Flush-mounted from the side