EasyManua.ls Logo

Endress+Hauser NRF81 - Page 201

Endress+Hauser NRF81
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Tankside Monitor NRF81 Operating menu
Endress+Hauser 201
HTG evaluation: dependence on measured level
To calculate the level or density by HTG with the required accuracy, P1 and P2 have to be
covered by a certain product level. To avoid a measurement with an insufficient accuracy,
the calculation will stop before the level reaches the position of the pressure sensor.
Two parameters are defined for this purpose:
Minimum level
This parameter defines the position below which no level is accepted. If the calculation
leads to Tank level < Minimum level, the value of Minimum level will be displayed
instead of the calculated value.
Safety distance
This parameter defines the minimum amount of product which must be present above
the pressure sensor P1 or P2 for the level or density calculation to take place.
The device always uses the bigger of these two values as the switch-over point for
the level calculation.
If HTG mode (→  206) is set to P1 only or P1 + P3, the density is not calculated
and the Manual upper density parameter is used instead.
Case 1: H
P1
< Minimum level < H
P2
A
B
C
3
P3
P2
P1
1
2
5
4
A0028718
1 P1 position (→  182)
2 Safety distance (→  208)
3 P1-2 distance (→  184)
4 Safety distance (→  208)
5 Minimum level (→  207)
Level L is in area Calculation method for ρ
P
Calculation method for L
A calculated from pressure calculated from pressure
B ρ
P
held calculated from pressure
C ρ
P
held L = Minimum level

Table of Contents

Other manuals for Endress+Hauser NRF81

Related product manuals