Mounting Flowmeter Proline 200 PROFINET with Ethernet-APL
12 Endress+Hauser
4.3 Turning the display module
The display module can be turned to optimize display readability and operability.
A0032238
1. Loosen the securing clamp of the electronics compartment cover using an Allen key.
2. Unscrew cover of the electronics compartment from the transmitter housing.
3. Optional: pull out the display module with a gentle rotational movement.
4. Turn the display module to the desired position: Max. 8 × 45 ° in each direction.
5. Without display module pulled out:
Allow display module to engage at desired position.
6. With display module pulled out:
Feed the cable into the gap between the housing and main electronics module and plug
the display module into the electronics compartment until it engages.
7. Reassemble the transmitter in the reverse order.
4.4 Transmitter post-installation check
The post-installation check must always be performed after the following tasks:
• Turning the transmitter housing
• Turning the display module
Is the device undamaged (visual inspection)?
Are the securing screw and securing clamp tightened securely?