Commissioning Flowmeter Proline 200 PROFINET with Ethernet-APL
42 Endress+Hauser
7. In the Reference temperature parameter, enter the fluid temperature associated with
the reference density.
8. In the Linear expansion coefficient parameter, enter the expansion coefficient of the
fluid.
9. In the Specific heat capacity parameter, enter the heat capacity of the fluid.
10. In the Dynamic viscosity parameter, enter the viscosity of the fluid.
8.6.3 Gas applications
For accurate mass or corrected volume measurement, it is recommended to use the
pressure-/temperature-compensated sensor version. If this sensor version is not
available, read in the pressure via the . If neither of these two options is possible, the
pressure can also be entered as a fixed value in the Fixed process pressure parameter.
Flow computer available only with the order code for "Sensor version", option "mass"
(integrated temperature measurement)" or option "mass (integrated pressure/
temperature measurement)".
Single gas
Combustion gas, e. g. methane CH
4
Select medium
Navigation:
Setup → Medium selection
1. Call up the Medium selection wizard.
2. In the Select medium parameter, select the Gas option.
3. In the Select gas type parameter, select the Single gas option.
4. In the Gas type parameter, select the Methane CH4 option.
Configuring medium properties
Navigation:
Setup → Advanced setup → Medium properties
5. Open the Medium properties submenu.
6. In the Reference combustion temperature parameter, enter the reference combustion
temperature of the medium.
Configuring medium properties
Navigation:
Setup → Advanced setup → Medium properties
7. Open the Medium properties submenu.
8. In the Reference combustion temperature parameter, enter the reference combustion
temperature of the medium.

Table of Contents