Flowmeter Proline 200 PROFINET with Ethernet-APL Electrical connection
Endress+Hauser 15
Wire cross-section
0.8 mm
2
(AWG 18)
Loop resistance (direct
current)
44 Ω/km
Characteristic impedance at
31.25 kHz
100 Ω ±20%
Attenuation constant at 39.0
kHz
3 dB/km
Capacitive asymmetry 2 nF/km
Envelope delay distortion
(7.9 to 39 kHz)
1.7 ms/km
Shield coverage 90 %
The following cable types are suitable for non-hazardous areas, for example:
Siemens 6XV1 830-5BH10
Belden 3076F
Kerpen CEL-PE/OSCR/PVC/FRLA FB-02YS(ST)YFL
Maximum overall cable length
The maximum network expansion depends on the type of protection and the cable
specifications. The overall cable length comprises the length of the main cable and the length
of all spurs >1 m (3.28 ft).
The maximum overall cable length for cable type A: 1 900 m (6 200 ft)
If repeaters are used, the maximum permitted overall cable length is doubled. A maximum of
three repeaters are permitted between the user and master.
Maximum spur length
The line between the distribution box and field device is described as a spur. In the case of
applications in non-hazardous areas, the max. length of a spur depends on the number of
spurs >1 m (3.28 ft):
Number of spurs Max. length per spur
1…12 120 m (400 ft)
13…14 90 m (300 ft)
15…18 60 m (200 ft)
19…24 30 m (100 ft)
25…32 1 m (3 ft)
Number of field devices
For systems designed in accordance with the Fieldbus Intrinsically Safe Concept (FISCO) with
EEx ia types of protection, the cable length is limited to a maximum of 1 000 m (3 300 ft). A
maximum of 32 users per segment in non-hazardous areas or a maximum of 10 users in