Installation Proline Promag P 100
14 Endress+Hauser
6 Installation
6.1
Installation conditions
No special measures such as supports are necessary. External forces are absorbed by the
construction of the device.
6.1.1 Mounting position
Mounting location
h ≥ 2 × DN
h
A0017061
Installation in down pipes
Install a siphon or a vent valve downstream of the sensor in down pipes whose length h
5 m (16.4 ft). This precaution is to avoid low pressure and the consequent risk of damage to
the measuring tube. This measure also prevents the system losing prime, which could cause
air pockets.
h
2
1
A0017064
4 Installation in a down pipe
1 Vent valve
2 Pipe siphon
h Length of down pipe