Proline Promag P 100 Commissioning
Endress+Hauser 43
Navigation
"Setup" menu → Device tag
Parameter overview with brief description
Parameters Description User entry Factory setting
Device tag Enter tag for measuring
point.
Max. 32 characters, such as
letters, numbers or special
characters (e.g. @, %, /).
Promag
10.6 Protecting settings from unauthorized access
The following options exist for protecting the configuration of the measuring device from
unintentional modification after commissioning:
Write protection via access code for Web browser (→  43)
Write protection via write protection switch (→  44)
10.6.1 Write protection via access code
With the customer-specific access code, access to the measuring device via the Web browser is
protected, as are the parameters for the measuring device configuration.
Navigation
"Setup" menu → Advanced setup → Administration → Def. access code
Structure of the submenu
Define access code
Define access code
Confirm access code
Defining the access code via the Web browser
1. Navigate to the "Enter access code" parameter.
2. Define a max. 4-digit numeric code as an access code.
3. Enter the access code again to confirm the code.
The Web browser switches to the login page.
If no action is performed for 10 minutes, the Web browser automatically returns to the
login page.
The user role with which the user is currently logged on via the Web browser is indicated
by the Access status tooling parameter. Navigation path: Operation
→ Access status
tooling