Installation Proline Promass 83
12 Endress+Hauser
2.3.3 Installing the wall-mount housing
"
Caution!
• Make sure that the ambient temperature does not exceed the permitted range.
• Always install the wall-mount housing in such a way that the cable entries point downwards.
Mounted directly on the wall
Pipe mounting
A0007542
1. Connection compartment
2. Securing screws M6 (max. ø 6.5 mm (0.25");
screw head max. ø 10.5 mm (0.4"))
3. Housing bores for securing screws
A0007543
"
Caution!
Danger of overheating! If the device is mounted on a
warm pipe, make sure that the housing temperature
does not exceed +60 °C (+140 °F) which is the
maximum temperature permitted.