5.5 Software settings
5.5.1 Device address PROFIBUS DP/PA, Modbus RS485
Has to be set for measuring devices with the following communication methods:
•PROFIBUS DP
Device address range 0 to 126, factory setting 126
• Modbus RS485
Device address range 1 to 247, factory setting 247
The device address can be configured via:
• Miniature switches  see section Hardware settings, "Device address PROFIBUS DP/PA,
Modbus RS485"  ä 17
• Local operation  see description below
!
Note!
The COMMISSIONING SETUP must be executed before setting the device address.
Calling the Communication Quick Setup
1.
F
Enter the function matrix (starting with measured value display)
2.
P
Select the group QUICK SETUP
F
Confirm selection
3.
N
Select the QUICK SETUP COMMUNICATION function
4. Intermediate step if configuration is blocked:
P
Enter the code 83 (confirm with
F
) and thus enable configuration
5.
P
Go to Communication Quick Setup
6.
P
Select YES
F
Confirm selection
7.
F
Start Communication Quick Setup
8. Configure the individual functions/settings:
– Via
P
-key, select option or enter number
– Via
F
-key, confirm entry and go to next function
– Via
Q
-key, return to Setup Commissioning function
(settings already made are retained)
All the available functions of the measuring device and their configuration options as well as
additional Quick Setups, if available, are described in detail in the "Description of Device
Functions" Operating Instructions. The related Operating Instructions can be found on the
CD-ROM.
The measuring device is ready for operation on completion of the Quick Setup.