Commissioning Proline Promass 83
26 Endress+Hauser
5.5.2 Device address EtherNet/IP network
Has to be set for measuring devices with the communication method:
• EtherNet/IP
The device address can be configured via:
• Miniature switches  see section Hardware settings,
"Device address EtherNet/IP network"  ä 19
• Webserver  see description below
Software addressing is performed in the "Network Configuration" menu of the Webserver. Both
the IP address for the EtherNet/IP network and the IP address for the Webserver can be
configured. The measuring device has the following default addresses when delivered:
Addresses in the range from 0 to 254 are permitted (the address 255 is reserved for the broadcast
address).
!
Note!
• Software addressing is disabled if hardware addressing is activated  ä 17.
• When changing from software addressing to hardware addressing, the first nine digits (first
three octets) that were configured using software addressing, remain unchanged.
• A reset of the software addressing to the default setting is possible  see SD00138D.
DHCP client
If a DHCP server is used within the EtherNet/IP network, the IP address, gateway and subnet
mask are set automatically when the DHCP client function of the Webserver is enabled. The
MAC address of the measuring device is used for identification purposes.
The DHCP client function is enabled in the "Network Configuration" menu.
The measuring device has the following DHCP default settings when delivered:
!
Note!
The DHCP client function is disabled if hardware addressing is enabled  ä 19.
EtherNet/IP network Webserver
IP address 192.168.212.212 192.168.212.213
Netmask 255.255.255.0 255.255.255.0
Gateway 192.168.212.212 192.168.212.213
EtherNet/IP network Webserver
DHCP Yes (enabled) No (disabled)