Proline Promass 83 Commissioning
Endress+Hauser
29
PROFIBUS
Modbus RS485
No. Device status message
(onsite display)
PROFIBUS measured value status Extended
diagnostic
message in
the
PROFIBUS
master
Cause/remedy
351
to
354
System error mess. (S)/
notice message (!)
CURRENT SPAN n
# 351 to 354
• Quality Code (HEX), measured
value status:
0x54; 0x55; 0x56
• Quality status:
UNCERTAIN
• Quality substatus:
Engineering unit range
violation
• Limits: O.K.; Low; High
Flow is out of
range
See HART table
701
Process error mess. (P)/
Notice message (!)
EXC. CURR. LIM
# 701
• Quality Code (HEX), measured
value status:
0x40; 0x41; 0x42
• Quality status:
UNCERTAIN
• Quality substatus:
Non-specific
• Limits: O.K.; Low; High
Excitation too
high
See HART table
Register 6859
Data type:
Integer
Register 6821
Data type:
String (18 byte)
No. Error message / Type Cause/remedy
39 to 42
RANGE
CUR. OUT n
351
to
354
System error message (S)/
Notice message (!)
CURRENT SPAN n
# 351 to 354
See HART table
112
EXC. CURR.
LIM
701
Process error mess. (P)/
Notice message (!)
EXC. CURR. LIM
# 701
See HART table