Installation Proline Prosonic Flow B 200 HART
20 Endress+Hauser
Two-path version: DN 100 to 200 (4 to 8")
10 × DN 3 × DN
1
3
2
4
10 × DN 3 × DN
10 × DN 3 × DN10 × DN 3 × DN
A0015553
5 Two-path version: minimum inlet and outlet runs with various flow obstructions
1 90 ° elbow or T-section
2 Pump
3 2 × 90 ° elbow, 3-dimensional
4 Control valve
Outlet runs when installing external devices
If installing an external device, observe the specified distance.
A0015901
PT Pressure transmitter
Installation dimensions
For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section
6.1.2 Requirements from environment and process
Ambient temperature range
Transmitter –40 to +60 °C (–40 to +140 °F)
Local display –20 to +60 °C (–4 to +140 °F), the readability of the display may be
impaired at temperatures outside the temperature range.
Sensor • Flange material carbon steel: –10 to +60 °C (+14 to +140 °F)
• Flange material stainless steel: –40 to +60 °C (–40 to +140 °F)
• Version without flange: –40 to +60 °C (–40 to +140 °F)
‣
If operating outdoors:
Avoid direct sunlight, particularly in warm climatic regions.
System pressure
Sensor
Max. 10 bar (145 psi)