Proline Prosonic Flow B 200 HART Installation
Endress+Hauser 19
Orientation Compact version
C Horizontal orientation, transmitter
head down *
A0015590
D Horizontal orientation, transmitter
head at side
A0015592
* A maximum deviation of only ±3° is permitted for the horizontal alignment of the
converters.
A0016534
Inlet and outlet runs
If possible, the sensor should be installed upstream from valves, T-pieces, elbows etc. To
attain the specified level of accuracy of the measuring device, the below mentioned inlet
and outlet runs must be maintained at minimum. If there are several flow disturbances
present, the longest specified inlet run must be maintained.
Single-path version: DN 50 (2"), DN 80 (3")
20 × DN 3 × DN
1
3
2
4
20 × DN 3 × DN
20 × DN 3 × DN20 × DN 3 × DN
A0015453
4 Single-path version: minimum inlet and outlet runs with various flow obstructions
1 90 ° elbow or T-section
2 Pump
3 2 × 90 ° elbow, 3-dimensional
4 Control valve