Proline Prosonic Flow B 200 HART Commissioning
Endress+Hauser 63
Parameter overview with brief description
Parameter Prerequsite Description Selection / User
entry
Factory setting
Select gas type Select measured gas type. Biogas
Coal seam gas
Air
Nitrogen N2
Natural gas
User-specific
biogas
Biogas
Methane fraction The following conditions are
met:
Order code for "Sensor
version", option 1 "Volume
flow"
In the Select gas type
parameter, the User-
specific biogas option is
selected.
If ordered without biogas
analysis feature, enter the
methan content of biogas.
Positive floating-
point number
55 %
Oxygen fraction In the Select gas type
parameter, the User-specific
biogas option is selected.
Enter O2 content of biogas to
reduce CH4 analysis
uncertainty.
Positive floating-
point number
0 %
Nitrogen fraction In the Select gas type
parameter, the User-specific
biogas option is selected.
Enter N2 content of biogas to
reduce CH4 analysis
uncertainty.
Positive floating-
point number
0 %
Additional gas component In the Select gas type
parameter, the User-specific
biogas option is selected.
Select additional gas
component of biogas to reduce
CH4 analysis uncertainty.
None
Hydrogen H2
Ammonia NH3
Hydrogen sulfide
H2S
None
Relative humidity In the Select gas type
parameter, the User-specific
biogas option is selected.
Enter humidity content of
biogas in %.
0 to 100 % 100 %
Relative humidity In the Select gas type
parameter, the Air option is
selected.
Enter humidity content of air
in %.
0 to 100 % 50 %
Pressure compensation Select pressure compensation
type.
Fixed value
External absolute
pressure
External gauge
pressure
Fixed value
Absolute pressure value In the Pressure compensation
parameter, the Fixed value
option is selected.
Enter process pressure to be
used for pressure correction.
700 to 11 000 mbar Country-specific:
1 043 mbar a
15.1 psi a
Atmospheric pressure In the Pressure compensation
parameter, the External
gauge pressure option is
selected.
Enter atmospheric pressure
value to be used for pressure
correction.
Dependency
The unit is taken from
the Pressure unit
parameter
700 to 1 100 mbar Country-specific:
1 013.25 mbar a
14.696 psi a
Process temperature For the following order code:
"Sensor version", option 1
"Volume flow"
Enter a fixed temperature
value for corrected volume
flow calculation.
0 to 80 °C Country-specific:
50 °C
122 °F
10.4.4 Configuring the current input
The "Current input" submenu guides you systematically through all the parameters that
have to be set for configuring the current input.