EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow B 200 - Page 77

Endress+Hauser Proline Prosonic Flow B 200
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Prosonic Flow B 200 HART Commissioning
Endress+Hauser 77
Assign
dir.check
End of wizard
Operating
mode
Switch
Switch
out funct
Energy flow
Mass flow
Volume flow
Failure mode
Invert outp.sig.
Corr CH4
volflow
Low flow cut
off
Assign
status
Fl. direct.
check
Status
Off
Correct.vol.
flow
A0019227-EN
 24 "Pulse/frequency/switch output" wizard in the "Setup" menu: "Operating mode" parameter"Switch" option
(part 3)
Parameter overview with brief description
Parameter Prerequsite Description Selection / User
entry
Factory setting
Operating mode Define the output as a pulse,
frequency or switch output.
Pulse
Frequency
Switch
Pulse
Switch output function The Switch option is selected
in the Operating mode
parameter.
Select function for switch
output.
Off
On
Diagnostic
behavior
Limit
Flow direction
check
Status
Off
Assign diagnostic behavior The Diagnostic behavior
option is selected in the Switch
output function parameter.
Select diagnostic behavior for
switch output.
Alarm
Alarm or warning
Warning
Alarm

Other manuals for Endress+Hauser Proline Prosonic Flow B 200

Related product manuals