Commissioning Proline Prosonic Flow B 200 HART
78 Endress+Hauser
Parameter Prerequsite Description Selection / User
entry
Factory setting
Assign limit In the Switch output function
parameter the Limit option is
selected.
Select process variable for limit
function.
Volume flow
Corrected volume
flow
Corrected methane
volume flow
*
Mass flow
Energy flow
*
Methane fraction
*
Wobbe index
*
Calorific value
*
Temperature
*
Totalizer 1
Totalizer 2
Totalizer 3
Sound velocity
Flow velocity
Acceptance rate
*
Signal asymmetry
*
Turbulence
*
Signal strength
*
Signal to noise
ratio
*
Volume flow
Assign flow direction check The Flow direction check
option is selected in the Switch
output function parameter.
Select process variable for flow
direction monitoring.
Off
Volume flow
Corrected volume
flow
Corrected methane
volume flow
Mass flow
Energy flow
Volume flow
Assign status The Status option is selected
in the Switch output function
parameter.
Select device status for switch
output.
Low flow cut off Low flow cut off
Switch-on value In the Switch output function
parameter the Limit option is
selected.
Enter measured value for the
switch-on point.
Signed floating-point
number
Country-specific:
0 m³/h
0 ft³/h
Switch-off value In the Switch output function
parameter the Limit option is
selected.
Enter measured value for the
switch-off point.
Signed floating-point
number
Country-specific:
0 m³/h
0 ft³/h
Switch-on delay The Limit option is selected in
the Switch output function
parameter.
Define delay for the switch-on
of status output.
0.0 to 100.0 s 0.0 s
Switch-off delay The Limit option is selected in
the Switch output function
parameter.
Define delay for the switch-off
of status output.
0.0 to 100.0 s 0.0 s
Failure mode Define output behavior in
alarm condition.
Actual status
Open
Closed
Open
Invert output signal Invert the output signal. No
Yes
No
* Visibility depends on order options or device settings
10.4.7 Configuring the local display
The Display wizard guides you systematically through all the parameters that can
configured for configuring the local display.