Installation Proline t-mass 65
14 Endress+Hauser
2.4 Post-installation check
• Is the measuring device damaged (visual inspection)?
• Does the measuring device correspond to the specifications at the measuring point?
• Are the measuring point number and labeling correct (visual inspection)?
• Correct internal diameter and correct surface roughness/quality?
• Correct alignment of the pipe/seal/body of the flow measuring device?
• Has the correct sensor orientation been selected in terms of type, fluid properties, fluid
temperature?
• Does the arrow on the sensor point in the direction of the flow in the pipe?
• Are sufficient inlet and outlet runs available before and after the measuring point?
• Flow conditioner correctly installed (if available)?
• Correct sensor immersion depth (only insertion version)?
• Is the measuring device protected against moisture and sunlight?
• Is the measuring device protected against overheating?
• Is the measuring device protected against excessive vibrations?
• Check gas properties (e.g. purity, dryness, cleanliness).