Commissioning Prosonic M - FOUNDATION Fieldbus
40 Endress+Hauser
6 Commissioning
This chapter consists of the following sections:
• 6.1 Function check ä 40
• 6.2 Unlocking the device ä 40
• 6.3 Parameter reset ä 42
• 6.4 Commissioning with display and operating module VU331 ä 44
• 6.5 Commissioning with FOUNDATION Fieldbus configuration tool ä 55
6.1 Function check
Make sure that all final checks have been completed before you start up your measuring point:
• Checklist “Post installation check” ( ä 26).
• Checklist “Post connection check” ( ä 30).
6.2 Unlocking the device
Before commissioning, make sure that the device is not locked against parameter changes. On
delivery, the device is unlocked. In other cases, however, it may have been locked in one of the
following ways:
6.2.1 DIP switch (under the housing cover)
Locking and unlocking
L00-FMU4xxxx-19-00-00-xx-002
1 Default settings: SIM = off (simulation); WP = off (write protection)
WP = on: parametrization locked
WP = off: parametrization unlocked
SIM = on: simulation possible in Analog Input Block by configuration tool
SIM = off: simulation not possible in Analog Input Block by configuration tool
Parameters affected
Locking by the DIP switch affects all parameters.
E
N
D
R
E
S
S
+
H
A
U
S
E
R
M
I
C
R
O
P
I
L
O
T
I
I
ENDRESS+HAUSER
MICROPILOT II
I
P
6
5
IP 65
O
r
d
e
r
C
o
d
e
:
S
e
r
.
-
N
o
.
:
Or
de
r
C
o
de:
Ser.-No.:
M
e
s
s
b
e
r
e
i
c
h
M
e
a
s
u
r
i
n
g
r
a
n
g
e
M
es
sber
e
i
c
h
Measuring range
U
1
6
.
.
.
3
6
V
D
C
4
.
.
.
2
0
m
A
U 16..
.
36V DC
4
.
.
.
20 mA
m
a
x
.
2
0
m
max
.
20 m
Made in Germany MaulburgMade in Germany Maulburg
T
>
7
0
°
C
:
A
t
>
8
5
°
C
T
>70°C :
A
t >
8
5
°C
on
off
SIM WP
SIM WP
off
on
1