Prosonic T Commissioning
Endress+Hauser 29
Function "process conditions" (004)
For this function, you have the following options:
standard liquids calm surface turb. surface
For all fluid applications which do not
fit in any of the following groups.
Storage tanks with immersion tube or
bottom filling
Storage / accumulation tanks with
uneven surface due to free filling,
mixing nozzles or small bottom stirrers
L00-FMU30xxx-14-00-00-xx-001 L00-FMU30xxx-14-00-00-xx-002
The filters and output damping are set
to average values.
The averaging filters and output
damping are set to large values.
-> Stable measured value
-> Accurate measurement
-> Slow reaction time
Special filters for stabilising the input
signal are activated.
-> Stable measured value
-> Medium reaction time
add. agitator fast change standard solid
Moving surfaces (poss. with vortex
formation) due to agitators
Rapid level change, particularly in
small tanks
For all bulk solids applications which
do not fit in any of the following
groups.
L00-FMU30xxx-14-00-00-xx-003 L00-FMU30xxx-14-00-00-xx-004 L00-FMU30xxx-14-00-00-xx-006
Special filters for stabilising the input
signal are set to large values.
-> Stable measured value
-> Medium reaction time
The averaging filters are set to small
values.
-> Rapid reaction time
-> Possibly unstable measured value
The filter and output damping are set
to average values.