Installation Prothermo NMT81
18 Endress+Hauser
The following tolerances are applied regardless of whether or not there is an optional WB
probe. The position of the flange cannot be adjusted in a welding flange type.
Probe length Tolerance of probe and element positions
1 000 to 25 000 mm (39.37 to 984.25 in) ± 50 mm (1.97 in)
25 001 to 40 000 mm (984.29 to 1 574.80 in) ± 50 mm (1.97 in)
40 001 to 60 000 mm (1 574.84 to 2 362.21 in) ± 100 mm (3.94 in)
60 001 to 100 000 mm (2 362.24 to 3 937.01 in) ± 300 mm (11.81 in)