Prothermo NMT81 Installation
Endress+Hauser 23
5.9 Pre-installation of NMT81
5.9.1 Unpacking
Unpack NMT81 with multiple people. If NMT81 is unpacked by one person alone, the
temperature probe may become bent or twisted.
A0042787
 14 Unpacking NMT81
5.9.2 Temperature probe handling
Do not pull the converter while holding onto the temperature probe. This may cause the
device to malfunction.
A0042788
 15 Temperature probe handling
When winding the temperature probe, keep the bending diameter at a minimum of
600 mm (23.62 in). When installing the temperature probe onto a tank or if it is
necessary to bend the temperature probe, ensure that the bent portion is at least R =
300 mm (11.81 in).