Installation Prothermo NMT81
24 Endress+Hauser
R
a
A0042789
 16 Installation and winding of a temperature probe
a 600 mm (23.62 in) or more
R 300 mm (11.81 in) or more
L
CAUTION
If bending the temperature probe with R smaller than 300 mm (11.81 in), it may
damage the probe and elements.
Bend the probe 300 mm (11.81 in) or more.
5.9.3 Installation height adjustment
A unique feature of NMT81 is the optional ability to adjust height by approx.
±180 mm (7.09 in) from the original position.
The height adjustment feature is not available for the welded flange type and the
converter-only version.
5.10 Installation procedure
The length of the NMT81 probe is predetermined by the customer. Check the following
items before installing.
Tag number on the device
Length of the temperature probe
Number of elements
Element intervals
The procedure for installing NMT81 will vary depending on the tank's shape and type. A
cone roof tank and a floating roof tank are used for the following examples. The
procedure to mount NMT81 flange on a tank nozzle flange is the same regardless of
what type of tank is being used.
The recommended diameter of the installation nozzle is:
Temperature only probe : 32A (1-1/4") or more
With WB probe: 50A (2") or more