Commissioning RIA15
36 Endress+Hauser
Setup menu (SETUP)
Parameters Values visible at Description
BGLO1-
BGLO4
Numerical value
–19 999 to 99 999
Default: 0.0
MODE =
HART
5-digit value (number of decimal places as configured
under DEC1-DEC4) for scaling the lower range of the bar
graph for HART1 - HART4.
The bar graph is disabled if BGLOx and BGHIx are set to
"0.0".
BGHI1-BGHI4 Numerical value
–19 999 to 99 999
Default: 0.0
MODE =
HART
5-digit value (number of decimal places as configured
under DEC1-DEC4) for scaling the upper range of the bar
graph for HART1 - HART4.
The bar graph is disabled if BGLOx and BGHIx are set to
"0.0".
UNIT1-UNIT4 HART
%
°C
°F
K
USER
MODE =
HART
Use this function to select the unit for displaying the
HART® value.
If "HART" is selected, the unit configured on the sensor/
actor is automatically adopted for the relevant HART®
value. Only units with a maximum of 5 characters can be
shown. Longer units are displayed as unit code "UCxxx".
The table in the HART® communication section at the
end of these Operating Instructions provides an overview
of the units that can be displayed.
If "USER" is selected, a user-defined unit can be entered in
the TEXT1-TEXT4 parameter.
TEXT1-TEXT4 Customized text,
5-digit
MODE =
HART
User-defined unit. Only visible if the "USER" option has
been selected under UNIT
Diagnostics menu (DIAG)
Parameters Values Description
AERR Read only The current diagnostic message appears on the display. If two or more
messages occur simultaneously, the message with the highest priority
is shown on the display.
LERR Read only The last diagnostic message with the highest priority appears on the
display.
FWVER Read only The firmware version appears on the display.
TERR
1)
Read only Displays the diagnostic code/error code pending at Endress+Hauser
HART® transmitters/sensors. Please refer to the Operating
Instructions of the relevant Endress+Hauser transmitter/sensor for
additional information about the meaning of the diagnostic number
and the remedial measures.
1) For Endress+Hauser transmitters/sensors with HART® communication, the diagnostic code/error code
currently pending can be queried via Endress+Hauser command #231. This command is only supported by
Endress+Hauser transmitters/sensors. Therefore, the TERR parameter is not visible if third-party devices
are connected to the RIA15.